LISUN Spectroradiometer for LED Testing

Table of Contents

Advanced Spectroradiometric Characterization of Solid-State Lighting with the LISUN LMS-6000 Series

Introduction to High-Fidelity Optical Radiation Measurement

The proliferation of solid-state lighting (SSL) technologies, primarily Light Emitting Diodes (LEDs) and Organic Light Emitting Diodes (OLEDs), has necessitated a paradigm shift in photometric and radiometric measurement methodologies. Unlike traditional incandescent sources, the spectral power distribution (SPD) of SSL is inherently narrowband and discontinuous, rendering traditional filter-based photometers inadequate for accurate characterization. Spectroradiometry, the science of measuring the absolute spectral power of a light source, has thus become the de facto standard for SSL testing. The LISUN LMS-6000 series of spectroradiometers represents a state-of-the-art instrument platform engineered to meet the rigorous demands of modern lighting development, quality control, and application-specific validation across a diverse range of industries. This article details the technical principles, specifications, and critical applications of this measurement system, with a specific focus on the LMS-6000S variant, optimized for high-sensitivity measurements.

Fundamental Operating Principles of Array Spectroradiometry

The core of the LMS-6000 series is based on a high-precision optical bench utilizing a diffraction grating and a charge-coupled device (CCD) or complementary metal-oxide-semiconductor (CMOS) detector array. The measurement process is governed by a sequence of physical and computational steps. Incident light from the source under test is collected via an integrating sphere or a fiber optic cable with a cosine corrector, which ensures angular response conformity to the Lambertian cosine law for illuminance measurements. This light is then directed into the spectrometer’s entrance slit.

Within the spectrometer, a collimating mirror directs the light onto a planar diffraction grating. This optical component angularly disperses the polychromatic light into its constituent wavelengths. The diffracted light is then focused by a second mirror onto the detector array. Each pixel on the array corresponds to a specific nanometer wavelength, allowing for the simultaneous capture of the entire SPD across the instrument’s operational range, typically 350-800nm for visible light applications or wider for specialized models. The raw signal from the detector is digitized and processed through a sophisticated software algorithm that applies crucial corrections for dark noise, stray light, and the system’s inherent spectral responsivity, which is calibrated traceably to national metrology institutes using standard lamps of known spectral output.

Technical Specifications of the LISUN LMS-6000S Spectroradiometer

The LMS-6000S is distinguished by its high sensitivity and low stray light characteristics, making it particularly suitable for measuring low-intensity sources, such as individual LED die, OLED panels, or displays in dark-room conditions. Its specifications are meticulously defined to comply with international standards such as CIE 63, CIE 177, and IES LM-79.

Table 1: Key Specifications of the LISUN LMS-6000S Spectroradiometer

Parameter Specification Significance
Wavelength Range 350 nm – 800 nm Covers the entire visible spectrum and parts of near-UV and near-IR relevant to photobiology and radiometry.
Wavelength Accuracy ± 0.3 nm Ensures precise identification of peak emission wavelengths, critical for color science and material analysis.
Wavelength Resolution ≤ 2.0 nm FWHM Provides sufficient detail to resolve narrow spectral features of phosphor-converted and RGB LEDs.
Dynamic Range 3,000,000:1 Allows for accurate measurement of very dim and very bright sources without changing hardware configurations.
Photometric Accuracy ± 2% (vs. NIM standard) Guarantees reliable measurements of luminous flux (lumens) and illuminance (lux).
Stray Light < 0.05% (at 600 nm) Minimizes spurious signal, which is crucial for measuring sources with deep spectral valleys, like white LEDs.
CCD Detector Cooling Thermoelectric cooling to -5°C Reduces dark current noise, enhancing signal-to-noise ratio (SNR) for low-light measurements.
Integration Time 1 ms to 20 s Provides flexibility to measure a vast range of source intensities, from a single micro-LED to a high-bay luminaire.

Critical Applications in LED and OLED Manufacturing

In the manufacturing environment, the LMS-6000S is integral to both Research & Development (R&D) and production line quality assurance (QA). During R&D, engineers utilize the instrument to characterize the efficacy (lumens per watt), color rendering index (CRI), correlated color temperature (CCT), and Duv of new LED architectures and phosphor formulations. The high wavelength accuracy is essential for ensuring that the chromaticity coordinates (x, y) fall within the target binning ellipse as defined by standards such as ANSI C78.377.

On the production floor, the LMS-6000S is deployed for high-speed spectral binning. By integrating the spectroradiometer with an automated test station, manufacturers can sort LEDs into precise chromaticity and flux bins at a rate of thousands of units per hour. This ensures consistency in the final lighting product and minimizes color shift between individual LEDs within a single luminaire. For OLED manufacturing, the instrument’s high sensitivity is critical for characterizing the uniformities and angular color stability of large-area, low-luminance panels.

Validation of Automotive Lighting Systems

Automotive lighting, encompassing headlamps, daytime running lights (DRLs), signal lights, and interior displays, is subject to stringent international regulations (e.g., ECE, SAE, FMVSS). The LMS-6000S, when coupled with a goniophotometer, is used to perform complete photometric and colorimetric analyses of these systems. Key measurements include the total luminous flux of a tail lamp, the precise chromaticity of a red stop lamp to ensure it meets the defined color fence, and the intensity distribution of a headlamp beam pattern. The ability to measure the SPD of adaptive driving beam (ADB) systems, which dynamically adjust their output, is paramount for safety and regulatory compliance.

Ensuring Performance in Aerospace and Aviation Lighting

The reliability requirements in aerospace are unparalleled. Aircraft navigation lights, anti-collision beacons, cockpit instrument backlighting, and passenger cabin lighting all require rigorous testing. The LMS-6000S can be used to verify that navigation lights maintain the mandated chromaticity and intensity over extreme temperature cycles and vibration profiles, as per standards like RTCA DO-160. Furthermore, the testing of lighting used in airport runways and signage demands exceptional accuracy to ensure pilot safety under all weather conditions.

Precision in Display and Photovoltaic Device Testing

In the display industry, the LMS-6000S is employed to calibrate and characterize the performance of LCD, OLED, and micro-LED displays. Measurements include screen uniformity, color gamut coverage (e.g., sRGB, DCI-P3, Rec. 2020), white point stability, and flicker percentage. The instrument’s fast sampling rate allows for temporal analysis of pulse-width modulation (PWM) dimming, which can cause visual discomfort.

Conversely, in the photovoltaic (PV) industry, the spectroradiometer is used not to measure light output, but to quantify the spectral irradiance of solar simulators. The accuracy of PV cell efficiency testing is directly dependent on the match between the simulator’s spectrum and the reference AM1.5G solar spectrum (defined by ASTM E927). The LMS-6000S provides the data necessary to classify a solar simulator (e.g., Class A, B, or C per IEC 60904-9) based on its spectral match.

Supporting Scientific Research and Medical Applications

In scientific laboratories, the LMS-6000S facilitates research in photobiology, horticulture, and material science. For instance, researchers studying the effects of light on plant growth can measure the precise SPD of horticultural LED fixtures to calculate photosynthetic photon flux density (PPFD) and photon efficacy. In medical lighting, the accurate measurement of surgical lights is critical, with parameters such as color rendering (particularly for tissue differentiation) and shadow dilution being key performance indicators. The instrument can also characterize the narrowband emission of phototherapy devices used for treating neonatal jaundice or skin conditions.

Advantages of the LMS-6000S in Demanding Measurement Scenarios

The competitive advantage of the LMS-6000S lies in its optimized design for high-fidelity, low-noise measurements. The thermoelectrically cooled detector is a significant differentiator, as it substantially lowers the noise floor. This enables accurate characterization of light sources that were previously challenging to measure, such as low-power OLEDs, dimmable LEDs at their minimum setting, and the dark-state of high-contrast displays. Its low stray light specification ensures that the measured values for color purity and chromaticity are not artificially skewed by out-of-band energy, a common issue with inferior spectrometers. This combination of high sensitivity and low stray light makes the LMS-6000S an indispensable tool for applications where measurement precision is non-negotiable.

Integration and Compliance with Global Standards

Deployment of the LMS-6000S is supported by comprehensive software that not only controls the instrument but also calculates over 100 photometric, radiometric, and colorimetric parameters. The system is designed for seamless integration with ancillary equipment such as integrating spheres, goniophotometers, and constant current sources. Compliance with standards such as LM-79-19 (Electrical and Photometric Measurements of Solid-State Lighting Products) and LM-80-20 (Measuring Lumen Maintenance of LED Light Sources) is built into the measurement workflows, ensuring that data generated is auditable and recognized within the global lighting industry.

Frequently Asked Questions (FAQ)

Q1: What is the primary distinction between the LMS-6000S and other models in the LMS-6000 series, such as the LMS-6000P?
The primary distinction lies in the detector optimization. The LMS-6000S incorporates a cooled CCD detector optimized for high sensitivity and low-noise measurements of low-intensity light sources. In contrast, the LMS-6000P typically features a high-linearity, non-cooled detector optimized for high-speed testing and high-intensity applications commonly found on production lines, trading some sensitivity for faster integration times and lower cost.

Q2: How does the instrument maintain calibration accuracy over time and environmental changes?
The LMS-6000S features robust housing and thermal management to minimize drift. However, regular calibration is essential for maintaining metrological traceability. The system should be recalibrated annually using NIST-traceable standard lamps. For critical applications, users can perform a quick wavelength calibration using known spectral lines from a mercury-argon lamp.

Q3: Can the LMS-6000S measure the flicker of LED lighting?
Yes, the instrument, when operated in its fast acquisition mode, can capture rapid changes in light output. The accompanying software can then analyze this temporal waveform to calculate flicker metrics such as percent flicker, flicker index, and frequency, as per standards like IEEE PAR1789.

Q4: What type of accessory is required to measure the luminous flux (lumens) of an LED lamp?
To measure total luminous flux, the spectroradiometer must be used in conjunction with an integrating sphere. The lamp is placed inside the sphere, and the spectroradiometer, connected via a fiber optic cable, measures the spatially averaged spectral radiance inside the sphere, which is then computationally converted to total luminous flux.

Q5: Is the system capable of measuring the optical characteristics of a light source at different angles?
The spectroradiometer itself measures light from a single direction. To perform angular measurements, it must be integrated with a mechanical goniophotometer. The goniophotometer rotates the light source or the detector, allowing the spectroradiometer to capture the SPD at numerous discrete angles, enabling the calculation of intensity distributions and angular color uniformity.