How to Measure Lumens with an Integrating Sphere

Table of Contents

Principles of Luminous Flux Measurement

Luminous flux, measured in lumens (lm), quantifies the total perceived power of light emitted by a source. It is a weighted integral of the radiometric power, where the spectral sensitivity of the human eye, defined by the CIE photopic luminosity function V(λ), is applied. Accurate measurement of this parameter is fundamental across industries, from ensuring consumer lighting products meet efficacy standards to guaranteeing the performance and safety of specialized illumination systems in aviation and medicine. The most accurate method for determining total luminous flux employs an integrating sphere, a device designed to create a spatially uniform radiance distribution from a light source under test (LUT).

The core principle of an integrating sphere is based on multiple diffuse reflections. The interior of the sphere is coated with a material exhibiting highly reflective and perfectly diffuse (Lambertian) properties. When light from the LUT is introduced into the sphere, it undergoes numerous reflections off the interior wall. With each reflection, the spatial information of the incident beam is scrambled. After several reflections, the irradiance on any point on the sphere’s wall becomes uniform and proportional to the total luminous flux entering the sphere. A detector, typically a spectroradiometer, is then placed at a port on the sphere to sample this uniform irradiance. By comparing the reading from the LUT to that of a standard lamp of known luminous flux, the absolute luminous flux of the LUT can be precisely calculated.

The Role of the Spectroradiometer in Photometric Quantification

While a photometer with a V(λ)-corrected filter can measure illuminance from which luminous flux is derived, a spectroradiometer provides a far more powerful and fundamental measurement. A spectroradiometer measures the absolute spectral power distribution (SPD) of the light source. The luminous flux Φ_v is then computed by integrating the SPD, S(λ), with the CIE V(λ) function across the visible wavelength range.

Φ_v = K_m ∫ S(λ) V(λ) dλ

Where K_m is the maximum spectral luminous efficacy, 683 lm/W. This method is inherently more accurate, especially for light sources with non-continuous spectra like Light Emitting Diodes (LEDs) or Organic LEDs (OLEDs), where the imperfect V(λ) matching of a filtered photometer can lead to significant errors. Spectroradiometric systems also yield a wealth of additional data, including chromaticity coordinates, correlated color temperature (CCT), and color rendering index (CRI), making them the instrument of choice for comprehensive light source characterization.

System Configuration: Integrating Sphere and Spectroradiometer Synergy

A complete luminous flux measurement system comprises three primary components: the integrating sphere, the spectroradiometer, and a calibrated power supply for the LUT. The sphere itself is a hollow spherical cavity with multiple ports. A main port is used for introducing the LUT, which is often mounted on a base that allows for electrical connection and thermal management. A second port accommodates the detector, which must not “see” the LUT directly but only the diffusely reflected light from the sphere wall. A third, smaller port may be used for an auxiliary lamp, which is critical for the self-absorption correction procedure.

The spectroradiometer is connected to the sphere via a fiber optic cable, which terminates at a cosine corrector attached to the detector port. This setup ensures the instrument captures light from a wide field of view, as required for accurate spatial integration. The entire system must be controlled by software that orchestrates the measurement sequence, from controlling the spectroradiometer’s integration time to calculating the final photometric and colorimetric values based on the acquired SPD.

The LPCE-3 High-Precision Spectroradiometer Integrating Sphere System

The LPCE-3 system represents an advanced implementation of this measurement methodology, designed to comply with the stringent requirements of LM-79-19, IESNA, and CIE 84 standards. It is engineered for the testing of single LEDs, LED modules, and complete luminaires. The system’s architecture is optimized for maximum accuracy and repeatability in both scientific research and high-volume manufacturing environments.

The core of the LPCE-3 is a modular integrating sphere available in various diameters (e.g., 0.5m, 1.0m, 1.5m, 2.0m) to accommodate light sources of different physical sizes and total flux outputs. The interior is coated with a proprietary, highly stable Spectraflect® or equivalent diffuse reflectance coating, which provides a reflectance factor greater than 94% across the visible spectrum and exhibits excellent temporal stability. The system employs the LMS-9000C or a similar high-performance CCD spectroradiometer, which features a wavelength range of 350-800 nm, a wavelength accuracy of ±0.3 nm, and a high dynamic range to accurately capture the SPD of both very dim and very bright sources.

A key feature of the LPCE-3 is its integrated auxiliary lamp system, which automates the self-absorption correction process. This correction is vital because the LUT, when placed inside the sphere, absorbs a portion of the light that would otherwise be reflected, leading to a measurement error. The auxiliary lamp method quantifies this absorption effect by taking measurements with the auxiliary lamp both with and without the LUT present inside the sphere. The software then uses this data to apply a correction factor, significantly improving measurement accuracy, particularly for large or dark-colored luminaires.

Calibration Protocols and Traceability

The accuracy of any photometric measurement system is contingent upon a rigorous calibration chain. The LPCE-3 system is calibrated using standard lamps of known total luminous flux, traceable to national metrology institutes such as NIST (USA) or PTB (Germany). The calibration procedure involves measuring the spectroradiometer’s response to the standard lamp under identical geometric conditions used for subsequent LUT measurements.

The system software stores the calibration coefficient, which relates the measured signal to the absolute spectral irradiance. This calibration must be periodically verified and renewed to account for potential degradation of the sphere’s coating or shifts in the detector’s sensitivity. The calibration process also characterizes the system’s spatial non-uniformity and polarization dependence, ensuring that these factors do not introduce significant measurement uncertainty.

Advanced Application: Self-Absorption Correction Methodology

Self-absorption is a critical systematic error in integrating sphere photometry. The physical presence of the LUT, its socket, and any supporting structures inside the sphere alters the sphere’s total reflectance. Since these objects are not perfectly reflective, they absorb a fraction of the light that impinges on them, effectively reducing the sphere’s multiplier constant. The magnitude of this error depends on the size, geometry, and color of the LUT relative to the sphere’s interior.

The LPCE-3 system employs the proven auxiliary lamp method to correct for this effect. The procedure is as follows:

  1. With the sphere empty, the auxiliary lamp is activated, and the spectroradiometer reading is recorded (Reading_1).
  2. The LUT is installed in the sphere (but not powered), and the auxiliary lamp is activated again. The spectroradiometer reading is recorded (Reading_2).
  3. The self-absorption correction factor, k, is calculated as k = Reading_1 / Reading_2.
  4. During actual testing, the LUT is powered, and its uncorrected luminous flux, Φ_uncorrected, is measured.
  5. The final, corrected luminous flux is calculated as Φ_corrected = k * Φ_uncorrected.

This method is automated within the LPCE-3’s software, ensuring that even complex luminaires can be measured with high accuracy.

Industry-Specific Use Cases and Requirements

The precision offered by systems like the LPCE-3 is indispensable across a diverse range of industries, each with its own unique set of standards and performance criteria.

LED & OLED Manufacturing: In mass production, consistency is paramount. The LPCE-3 is used for binning LEDs based on luminous flux and chromaticity, ensuring that components destined for a single product fall within tight tolerances. For OLED panels used in displays or lighting, the system verifies uniformity and total light output against design specifications.

Automotive Lighting Testing: Automotive forward lighting (headlamps) and signal lighting (tail lights) are subject to rigorous international regulations (e.g., ECE, SAE). The LPCE-3 system can measure the total luminous flux of individual LEDs within a headlamp assembly or the entire signal lamp, ensuring compliance with minimum and maximum intensity requirements for safety.

Aerospace and Aviation Lighting: Cockpit displays, panel backlighting, and exterior navigation lights must perform reliably under extreme environmental conditions. The LPCE-3 provides the baseline photometric data used to qualify components, and its spectroradiometric capabilities are essential for ensuring that colors meet specific aviation standards for readability and recognition.

Display Equipment Testing: For LCD, OLED, and microLED displays, the LPCE-3 can be used to measure the luminous flux of backlight units (BLUs) and the overall screen’s luminance and color gamut. This is critical for quality control in the production of televisions, monitors, and mobile device screens.

Photovoltaic Industry: While primarily concerned with energy conversion, the photovoltaic industry uses high-power, spectrally tunable LED-based solar simulators to test solar cells. The LPCE-3 can be used to calibrate and verify the spectral irradiance and spatial uniformity of these simulators against defined standards like IEC 60904-9.

Medical Lighting Equipment: Surgical lights and dermatological treatment devices have strict requirements for intensity and color rendering to ensure accurate tissue differentiation and effective treatment. The LPCE-3 system provides the comprehensive spectral data necessary to validate that these medical devices meet their design and regulatory specifications.

Comparative Analysis of System Performance

The performance of an integrating sphere system is quantified by its measurement uncertainty and repeatability. Key factors influencing performance include sphere diameter, coating reflectance, and detector performance. The LPCE-3 system, with its high-reflectance coating and high-sensitivity spectroradiometer, is designed to minimize these uncertainties.

Table 1: Typical LPCE-3 System Specifications and Performance Metrics
| Parameter | Specification | Impact on Measurement |
| :— | :— | :— |
| Sphere Diameter | 0.5 m to 2.0 m | Larger spheres reduce spatial non-uniformity and self-absorption errors for bigger luminaires. |
| Coating Reflectance | >94% (350-800nm) | Higher reflectance increases signal-to-noise ratio and improves spatial integration. |
| Spectroradiometer Wavelength Range | 350-800 nm | Ensures full capture of the visible spectrum for accurate photopic and colorimetric calculations. |
| Luminous Flux Uncertainty | <3% (for standard LEDs) | A measure of the system's overall accuracy, traceable to national standards. |
| Repeatability | >99.5% | Indicates the system’s consistency in producing the same result under unchanged conditions. |

The primary competitive advantage of a system like the LPCE-3 lies in its integrated, automated approach to self-absorption correction and its use of a spectroradiometer as the primary detector. This eliminates the errors associated with V(λ) filter mismatch, a significant source of inaccuracy when measuring modern solid-state lighting sources with narrow-band or spiky spectral emissions. Furthermore, the system’s software provides not only luminous flux but a complete photometric and colorimetric report, increasing testing efficiency.

Mitigating Measurement Uncertainty and Error Sources

Achieving high accuracy requires a thorough understanding and mitigation of potential error sources. Key uncertainties in integrating sphere photometry include:

  1. Calibration Uncertainty: The inherent uncertainty of the standard lamp used for calibration propagates directly into all subsequent measurements.
  2. Self-Absorption Error: As discussed, this is a major error source for large or dark LUTs, but it is effectively corrected by the auxiliary lamp method in the LPCE-3.
  3. Spatial Non-uniformity: Imperfections in the sphere’s coating or geometry can lead to areas of higher or lower reflectance. A high-quality sphere minimizes this, and the detector’s position is optimized to sample a representative area.
  4. Stray Light: Light that enters the detector without undergoing sufficient reflections can cause errors. Proper baffling between the LUT port and the detector port is essential.
  5. Temperature Dependence: The output of many light sources, particularly LEDs, is sensitive to junction temperature. The LPCE-3 system incorporates temperature-stabilized mounts and requires the LUT to be measured under thermal steady-state conditions as defined by relevant standards.

By systematically addressing these factors through robust system design, precise calibration, and standardized operating procedures, the LPCE-3 maintains a low overall measurement uncertainty, making it suitable for both R&D and compliance testing.

Frequently Asked Questions

What is the difference between measuring lumens with a photometer versus a spectroradiometer inside an integrating sphere?
A photometer uses a filtered silicon photodiode that is corrected to match the CIE V(λ) function. However, this correction is never perfect, leading to “spectral mismatch errors,” especially for non-continuous light sources like LEDs. A spectroradiometer measures the complete spectral power distribution (SPD). The luminous flux is then calculated by mathematically integrating the SPD with the ideal V(λ) function, resulting in a fundamentally more accurate and versatile measurement.

Why is self-absorption correction necessary, and how does the LPCE-3 perform it?
Any object placed inside the integrating sphere, including the light source itself, its socket, and wiring, will absorb some light, reducing the overall measured signal. This leads to an underestimation of the true luminous flux. The LPCE-3 uses an integrated auxiliary lamp to quantify this absorption. By comparing the sphere’s response to the auxiliary lamp with and without the device under test present, the system calculates a precise correction factor that is automatically applied to the measurement.

For a large, complex luminaire, what size integrating sphere is recommended?
The sphere should be large enough so that the luminaire does not occupy more than a small fraction (typically recommended to be less than 5%) of the sphere’s total internal surface area. This minimizes self-absorption errors and preserves the spatial uniformity of the radiance inside the sphere. For a typical streetlight or large panel light, a 1.5m or 2.0m diameter sphere is often required. The LPCE-3 is available in these sizes to accommodate such applications.

Can the LPCE-3 system measure the luminous flux of a light source that emits primarily in the ultraviolet or infrared spectrum?
The standard LPCE-3 system is calibrated for the visible spectrum (typically 350-800 nm) for photopic (human vision) measurements. For radiometric measurements in the UV or IR ranges, the system would require a different spectroradiometer detector (e.g., with an InGaAs array for IR) and a sphere coating that is highly reflective in those specific wavelength regions. Such configurations are available as custom options.